| Pipeline gas input | O2/N2O and supplement O2 supply, input |
|---|---|
| pressure | 0.28MPa~0.6MPa |
| Cylinder gas input (optional) | Pin indexed yokes for O2, N2O |
| Pressure gauge | 2 nos. for pipeline. 2 nos. for cylinder (Optional) O2 controls |
| Method | Proportionate decrease of N2O with reduction in O2 flow |
| O2 supply failure alarm | O2 pressure lower than 0.14MPa |
| N2O cut off | N2O cutoff with loss of O2 pressure, supply failure alarm when <200KPa |
| O2 flush | 25~75L/min |
| Flowmeters | Four tube flow meter |
| Operational modes | Closed and Semi-closed circuit system |
| Hypoxic guard system | Mechanical type, provides a nominal minimum 25% concentration of oxygen in O2/N2O mixture |
| CO2 absorbent canister | with volume 2L |
| Bag/ventilation switch | switch for manual/spontaneous ventilation and mechanical ventilation |
| Integrated Adjustable Pressure Limiting (APL) valve | ≤ 6KPa |
| AGSS (optional) | Passive Anesthetic Gas Scavenging System |