Pipeline gas input O2/N2O/AIR and supplement O2 supply, input
pressure 0.28MPa~0.6MPa
Cylinder gas input (optional) Pin indexed yokes for O2, N2O
Pressure gauge 3 nos. for pipeline. 2 nos. for cylinder (Optional)
Method Proportionate decrease of N2O with reduction in O2 flow
O2 supply failure alarm O2 pressure lower than 0.14MPa
N2O cut off N2O cutoff with loss of O2 pressure, supply failure alarm when <200KPa
O2 flush 25~75L/min
Flowmeters Six tube flowmeter
Operational modes Closed and Semi-closed circuit system
Hypoxic guard system Mechanical type, provides a nominal minimum 25% concentration of oxygen in O2/N2O mixture
ACGO ISO 22mm OD and 15mm ID,allows for using third party's semi-open circuit
CO2 absorbent canister with volume 1.5L*2
Bag/ventilation switch switch for manual/spontaneous ventilation and mechanical ventilation
Integrated Adjustable Pressure Limiting (APL) valve ≤ 6KPa
AGSS (optional) Passive Anesthetic Gas Scavenging System